A&D FC-50Ki fine mode gives the 50 kg model a much smaller counting unit-weight threshold than its 5 g direct weighing division, but it should be used with a clear understanding of what the specification means. A&D lists 1 g minimum unit weight in normal mode and 0.05 g in fine mode, while the normal weight display resolves 5 g.

Fine mode is not 0.05 g direct readability

Fine mode leverages the FC-i family’s high internal resolution to estimate average piece weight for counting. The display is not converted into a laboratory balance. If you place a single 0.05 g object on an empty 50 kg scale, the normal weighing display is not intended to resolve that object as a direct measurement. Fine counting becomes meaningful when a representative sample produces enough total signal for the internal algorithm to estimate unit weight.

ACAI and sample discipline

ACAI improves the average piece weight after the initial sample. For high-count bins, this is important because a small unit-weight error multiplies across thousands of pieces. Start with clean parts from the actual lot, use a larger sample when piece mass varies, let the display stabilize and add material gradually enough for the algorithm to update. When lots change vendor, revision or material condition, rebuild the unit weight rather than blindly recalling an old value.

RS-232C planning

The FC-i family includes one standard RS-232C channel. FC-03i can provide two additional serial channels, while FC-04i provides one additional RS-232C channel plus comparator relay. A&D notes that FC-03i and FC-04i cannot be used together. A station needing a PC, printer and automation relay should therefore be planned around the actual destination signals before options are ordered.

Remote scale and capacity expansion

FC-05i enables a remote scale for two-scale counting. This is the preferred expansion route when a larger platform or floor-mounted container is needed. The remote load receiving element and indicator setup must be matched to the application. The FC-50Ki’s stored unit weights can support the workflow, but the remote scale’s own divisions and stability determine how much uncertainty enters the bulk count.

Calibration, verification and maintenance

MNM lists 50 kg as the FC-50Ki calibration weight. Calibration should be followed by verification at multiple loads and a known-count check using production parts. Inspect the stainless platform for contact with guards or debris, keep the bench rigid, avoid airflow from dock doors during sampling, and confirm that feet remain stable. If counts drift while known test weights remain correct, investigate the parts, sample procedure and container tare before assuming the load cell is faulty.

Optional battery

FC-02i is the rechargeable Ni-MH battery option. A&D lists approximately ten hours of typical operation and about fifteen hours for charging. A mobile FC-50Ki still needs a stable surface; the convenience of battery operation does not eliminate the need for leveling and vibration control.

 
 
 