How to Calibrate & Troubleshoot the A&D EK-300i
Need more details or help choosing the correct item? Call MNM Scales at (832) 290-3120 or email mnmscales@yahoo.com.
A&D EK-300i calibration uses an external reference mass; MNM lists 300 g as the factory/default calibration value. The EK/EW-i manual recommends leveling, thermal equilibrium and at least about 30 minutes of powered warm-up before precision weighing. Manufacturer performance is approximately 0.01 g repeatability and ±0.02 g linearity. A useful troubleshooting program verifies those behaviors under controlled conditions before changing calibration, because drafts, static, temperature and pan contact can easily move a hundredth-gram display.
Warm up and level first
Use the leveling feet/bubble and a rigid bench. Keep the balance away from fans, HVAC outlets, direct sun and vibrating equipment. Let a balance or calibration mass moved from a different temperature acclimate. Power on for at least about 30 minutes. A calibration performed while the load cell, electronics or reference mass is still warming may look correct briefly and drift afterwards.
External calibration with 300 g
With the pan clean/empty and stable, enter the calibration routine and use an appropriate known 300 g mass. Center it on the 110 mm pan, avoid sliding/touching the bench and wait for stability. Use a reference whose uncertainty fits the process. Do not substitute an unverified household object or another ordinary scale when calibration traceability matters.
Verify after calibration
After CAL, remove the mass and confirm zero, then test an independent known weight. If the process spans a broad fraction of 300 g, check more than one point. Repeat a key weight several times to assess repeatability. A completed calibration adjusts span; verification demonstrates whether the full setup meets the acceptance criteria.
Linearity and repeatability
A&D lists 0.01 g repeatability and ±0.02 g linearity. If repeated measurements of one known weight spread more than expected, investigate environment/mechanics before recalibrating. If error changes consistently with load—for example correct near 300 g but off near 50–100 g—perform multi-point verification and arrange service if necessary rather than repeatedly adjusting only the full-scale span.
Static and airflow
Plastic containers, films and powders can become electrostatically charged. At 0.01 g readability, static forces and convection around warm/cold samples can exceed a display division. If the reading changes when a hand approaches a container or when a fan cycles, improve static/airflow control. Do not calibrate around a changing environmental force.
Pan position and mechanical contact
Make sure the pan and any container touch only the weighing support. For a practical eccentric-load check, compare the same test weight in the center and several off-center positions. Large position-dependent errors indicate mechanical interference, leveling or damage—not a simple span issue. A center calibration cannot correct a container that touches the housing or bench.
RS-232C troubleshooting is separate from calibration
If the local display passes a known-weight check but a PC receives nothing/garbled data, troubleshoot cable pinout, baud/parity/data settings and output mode. Do not recalibrate a balance to fix a serial communications problem. Conversely, if both display and transmitted result fail verification, investigate the weighing system first.
Non-NTEP status does not change with calibration
EK-300i is not identified as an NTEP model in A&D’s current U.S. EK-i catalog. Calibration can improve/restore internal measurement performance, but it cannot turn the model into a certified commercial scale. For customer sale by weight, use an explicitly approved configuration.
Frequently asked questions
What is the default calibration mass? 300 g. How long should EK-300i warm up? At least about 30 minutes. What are repeatability/linearity? 0.01 g and ±0.02 g. Should calibration be followed by verification? Yes. What commonly causes last-digit instability? Drafts, static, vibration, temperature and pan contact. What if serial data is garbled but the local display is correct? Troubleshoot RS-232C, not calibration. Does calibration make EK-300i NTEP? No.