How to Calibrate, Verify & Integrate the A&D HC-30Ki

Jul 31
2 min read

Need more details or help choosing the correct item? Call MNM Scales at (832) 290-3120 or email mnmscales@yahoo.com.

A&D HC-30Ki calibration and verification starts with a 30 kg-capable stable station. A&D specifies a 30 kg ±1 g calibration weight, 5 g repeatability and ±5 g non-linearity. After the basic weight system is checked, verify a known piece count because incorrect sampling or stale memory can create inventory error even when the platform weighs correctly.

30 kg calibration and platform checks

Level the scale on a rigid bench and center a suitable 30 kg reference. Heavy containers can flex tables or touch adjacent structures, so inspect the real loading arrangement. After calibration, verify with another known load. Compare representative pan positions if operators often place bins off center; a mechanical corner-loading issue should be repaired rather than calibrated around.

Known-count verification

Recall or create the SKU unit weight, allow ACAI to update it and test a known hand-counted lot. If weight checks pass but count is wrong, inspect sample size, piece variability, tare and memory selection. Recalibration does not fix a wrong unit weight or the wrong SKU ID.

Data and relay integration

HC-03i provides RS-232C and HC-04i provides serial plus HI/OK/LO relay output. Validate local weight, local count, selected ID and transmitted record together. WinCT-UFC can help implement custom data formats. If software inventory disagrees with the local scale, trace the record path before touching calibration.

Battery and relocation

The optional SLA battery supports roughly 80 hours without interface. After moving the scale, re-level and verify. If readings differ on AC versus battery, troubleshoot power before assuming metrological drift. Inspect the pan for debris and surrounding contact whenever zero fails to return after a heavy bin is removed.

Frequently asked questions

What calibration weight is specified? 30 kg ±1 g. What are repeatability/non-linearity? 5 g and ±5 g. Should I also verify a known count? Yes. What if weight passes but count fails? Check sample, ACAI, tare and SKU memory. Does HC-04i provide relay outputs? Yes. Battery runtime? About 80 hours without interface. What if corner loading changes the result? Correct the mechanical setup. Is HC-30Ki legal for trade? No.

 
 
 