How to Calibrate the MNM Scales 20,000 lb x 2 lb Load Cell Scale

Jul 29
2 min read

For calibration help or to confirm the indicator supplied with your MS720-20K, call (832) 290-3120 or email mnmscales@yahoo.com.

To calibrate an MNM Scales 20,000 lb x 2 lb load cell scale correctly, start with the mechanical installation before touching any calibration menu. MNM sells the system calibrated, but recalibration may be needed after replacing the load cell or indicator, changing fixtures, repairing wiring, moving the system to a new application, or finding a consistent error during verification.

1. Confirm the Exact Indicator

MNM’s MS720-20K listing describes the indicator’s functions but does not lock the product page to one model number. That matters because service-menu key sequences vary. Read the model label on the indicator and use that model’s instructions rather than assuming an LP7515 or OP-901 sequence applies.

2. Inspect the Mechanical Setup

Remove unintended preload, make sure the S-beam hangs or mounts straight, inspect eye bolts and adapters, and verify that nothing is rubbing or binding. Calibration cannot compensate reliably for a fixture that side-loads the sensor. Let the load cell and indicator reach a stable temperature before precision verification.

3. Establish Zero

With the test path unloaded except for the normal dead-load hardware that will remain during use, establish the zero point according to the indicator manual. If a tare or fixture weight will always remain installed, decide whether the application should calibrate with that fixture present or whether the indicator should be calibrated at true unloaded sensor zero and then tared during operation.

4. Apply a Known Calibration Load

Use a traceable test load or force standard appropriate to the required uncertainty. The calibration point should be high enough in the working range to produce a meaningful span without exceeding rated capacity. Enter the known value through the indicator’s documented span/loading procedure and allow the load to stabilize before accepting the point.

5. Verify More Than One Point

Do not stop after one successful calibration point. Return to zero, then check several increasing loads across the range and, when practical, check decreasing loads as well. Watch for repeatability, return-to-zero behavior and a consistent bias. A system that is right at one point but wrong elsewhere may have a mechanical problem, damaged sensor, incorrect capacity/division setting or an unsuitable calibration procedure.

Representative 20,000 lb S-type sensor system for zero and span calibration

Common Calibration Problems

A reading that changes when the cable moves points toward wiring or connector trouble. A reading that changes when the fixture angle changes points toward side load or binding. Slow drift can come from temperature, creep, unstable mounting or power issues. Large nonlinear errors can indicate overload damage or a configuration mismatch. Always correct mechanical and electrical faults before repeatedly recalibrating.

Technical Reference

FAQ

How often should a 20,000 lb load cell scale be calibrated?

The interval depends on use, environment, quality requirements and applicable procedures. Verify more often after overload events, repairs, relocation, fixture changes or when readings are safety- or quality-critical.

Can I calibrate with a small weight?

A very small load relative to 20,000 lb full scale may not provide enough span information for a high-capacity system. Use a known load appropriate to the working range and the uncertainty you need.

 
 
 