top of page
A&D EK-3000EP intrinsically safe balance, angled front view

A&D EK-3000EP Intrinsically Safe, 3000 g × 0.1 g

SKU: 3I-4IRV-21OD

The A&D EK-3000EP intrinsically safe compact balance provides 3000 g × 0.1 g for portable weighing and process checks. Select it only when the supplied unit’s approval markings and the safety conditions below match your installation.

  • Specifications

    Brand A&D Weighing
    Model EK-3000EP
    Capacity × readability 3000 g × 0.1 g
    Repeatability (standard deviation) 0.1 g
    Linearity ±0.2 g
    Weighing pan 133 × 170 mm, stainless steel
    Display Backlit LCD, 16 mm characters; approximately 10 updates/second
    Weighing units g, oz, lb, ozt, mom and dwt; grain also available on this model. Regional tola/tael units are version-dependent.
    Functions Parts counting with ACAI, percent weighing, HI/OK/LO comparison, adjustable response and automatic power-off
    Counting samples 5, 10, 25, 50 or 100 pieces
    Minimum unit mass 0.1 g
    Maximum count at factory Umin setting 30,000 pieces
    Percent display increment 0.1%
    Minimum 100% reference weight 10 g
    Calibration External; factory calibration mass 3000 g; follow the manual for supported calibration masses
    Data interface No data-output interface specified in the EK-EP manual
    Power Four approved 1.5 V AA alkaline batteries; 6 V DC nominal; battery-only operation
    Battery runtime Approximately 250 hours with backlight off; varies with conditions and use
    Operating environment 0°C to 40°C; below 85% relative humidity, non-condensing
    Net weight Approximately 1.5 kg, excluding batteries
    Security Security ring for a separately supplied lock cable
  • What's Included & Options

    Includes the EK-3000EP balance with stainless-steel pan, instruction manual and 2.5 mm Allen wrench.

    Sold separately: approved AA alkaline batteries, calibration weights and the EJ-12 carrying case.

    Product photos show EK-3000EP; regional markings may vary. Photos do not establish the supplied unit’s safety certification.

  • Safety & Approved Power

    The manufacturer-hosted FM 3054278 US certificate and EK-EP manual identify Class I, Division 1, Groups C and D, T3C, and Class I, Zone 0, AEx ia IIB, T3. The manual specifies a −25°C to +40°C approval ambient range; the separate operating range is listed in Specifications. Match the actual unit’s markings, location classification and every certificate/manual condition before use.

    • Use only four Duracell MN1500 LR6, Energizer E91 LR6 AM3 or Panasonic LR6(XJ) AA alkaline batteries. All four must be the same type and manufacturer; do not mix old and new cells.
    • Install or replace batteries only in a non-hazardous area. Do not use rechargeable batteries or an AC adapter.
    • Assess the certificate’s 170 pF accessible-metal capacitance condition when determining electrostatic suitability; clean plastic surfaces only with a damp cloth.
    • Do not disassemble or modify the balance. Keep it away from corrosive gases and prevent it from getting wet or soaked.
  • Contact Us

    Questions on compatibility, calibration, or what is in the box? All Scales USA, 21249 FM 529 Rd, Cypress, TX 77433. Call (832) 766-5410 or email pete@allscales.com. Also: (281) 656-2666 · (832) 290-3120.

  • Model

    EK-3000EP
  • Product Type

    Intrinsically Safe Compact Balances
$1,245.23Price
Quantity
bottom of page