A&D EK-3000EP Intrinsically Safe, 3000 g × 0.1 g
The A&D EK-3000EP intrinsically safe compact balance provides 3000 g × 0.1 g for portable weighing and process checks. Select it only when the supplied unit’s approval markings and the safety conditions below match your installation.
Specifications
Brand A&D Weighing Model EK-3000EP Capacity × readability 3000 g × 0.1 g Repeatability (standard deviation) 0.1 g Linearity ±0.2 g Weighing pan 133 × 170 mm, stainless steel Display Backlit LCD, 16 mm characters; approximately 10 updates/second Weighing units g, oz, lb, ozt, mom and dwt; grain also available on this model. Regional tola/tael units are version-dependent. Functions Parts counting with ACAI, percent weighing, HI/OK/LO comparison, adjustable response and automatic power-off Counting samples 5, 10, 25, 50 or 100 pieces Minimum unit mass 0.1 g Maximum count at factory Umin setting 30,000 pieces Percent display increment 0.1% Minimum 100% reference weight 10 g Calibration External; factory calibration mass 3000 g; follow the manual for supported calibration masses Data interface No data-output interface specified in the EK-EP manual Power Four approved 1.5 V AA alkaline batteries; 6 V DC nominal; battery-only operation Battery runtime Approximately 250 hours with backlight off; varies with conditions and use Operating environment 0°C to 40°C; below 85% relative humidity, non-condensing Net weight Approximately 1.5 kg, excluding batteries Security Security ring for a separately supplied lock cable What's Included & Options
Includes the EK-3000EP balance with stainless-steel pan, instruction manual and 2.5 mm Allen wrench.
Sold separately: approved AA alkaline batteries, calibration weights and the EJ-12 carrying case.
Product photos show EK-3000EP; regional markings may vary. Photos do not establish the supplied unit’s safety certification.
Safety & Approved Power
The manufacturer-hosted FM 3054278 US certificate and EK-EP manual identify Class I, Division 1, Groups C and D, T3C, and Class I, Zone 0, AEx ia IIB, T3. The manual specifies a −25°C to +40°C approval ambient range; the separate operating range is listed in Specifications. Match the actual unit’s markings, location classification and every certificate/manual condition before use.
- Use only four Duracell MN1500 LR6, Energizer E91 LR6 AM3 or Panasonic LR6(XJ) AA alkaline batteries. All four must be the same type and manufacturer; do not mix old and new cells.
- Install or replace batteries only in a non-hazardous area. Do not use rechargeable batteries or an AC adapter.
- Assess the certificate’s 170 pF accessible-metal capacitance condition when determining electrostatic suitability; clean plastic surfaces only with a damp cloth.
- Do not disassemble or modify the balance. Keep it away from corrosive gases and prevent it from getting wet or soaked.
Contact Us
Questions on compatibility, calibration, or what is in the box? All Scales USA, 21249 FM 529 Rd, Cypress, TX 77433. Call (832) 766-5410 or email pete@allscales.com. Also: (281) 656-2666 · (832) 290-3120.
Model
EK-3000EPProduct Type
Intrinsically Safe Compact Balances