Moisture Analyzers
Compare moisture analyzers from A&D, Radwag and Sartorius by sample capacity, weight and moisture readability, heating system and reporting needs. Choose an instrument around a validated loss-on-drying method and safe sample compatibility. Review the models below, then confirm the exact configuration, accessories and availability with All Scales USA.
Start with the test method and sample requirements
A moisture analyzer combines a weighing system with controlled heating to measure loss on drying. Before choosing one, identify your approved test method, required result units, reporting needs and sample suitability. Maximum weighing capacity is an instrument limit, not a recommended test portion. Moisture percentage readability and weight readability are separate specifications; neither alone establishes the accuracy of a method.
Compare useful differences between models
The A&D MF-50 has a 51 g sample capacity, 0.002 g weight readability and moisture display increments down to 0.05%. For finer displayed increments, the A&D MS-70 provides 71 g capacity, 0.0001 g weight readability and moisture display down to 0.001%. Both use halogen heating. Select the resolution your validated method needs rather than assuming extra digits guarantee a better result.
The Radwag MA 110.X2.A.WH combines 110 g capacity and 1 mg weight readability with a halogen heater, automatic chamber and external weight adjustment. The Sartorius MA160 provides 200 g capacity, 1 mg weight readability and 0.01% moisture readability with AURI infrared heating. Its method-development assistant and storage for 100 methods are useful distinctions when several approved tests share one instrument.
Specify how tests finish and results are recorded
Choose endpoint controls that support your procedure. The MA 110.X2.A.WH offers automatic, manual, timed and user-defined finish modes. The MA160 offers fully automatic, semiautomatic, timed and manual shutoff. A timer and an automatic mass-stability endpoint answer different method requirements; confirm the required option before purchasing.
For documentation, the MF-50 and MS-70 provide RS-232C output. The Radwag model offers drying-program and report databases; the MA160 supports method and result export by SD card. Confirm the receiving software, printer, cable and report fields rather than assuming every connection supplies the same records.
Confirm safety, accessories and checks
Follow the exact model's safety instructions. Do not heat unknown, flammable or explosive samples, or materials that release hazardous substances. An instrument's temperature range does not establish sample suitability. Ask about compatible pans, the correct supply voltage, weight checks and separate temperature verification equipment. Calibration weights address the weighing system; they do not verify heater temperature. For ordinary weighing without a drying stage, explore lab and analytical equipment. Call (832) 766-5410 with your method and documentation requirements.