MNM Scales 4 x 5,000 lb Stainless NTEP Shear Beam Kit Guide: Wiring, Feet & Interchanges
For help confirming capacity, interchange compatibility or the correct indicator/junction-box setup, call MNM Scales at (832) 290-3120 or email mnmscales@yahoo.com.
The MNM Scales four 5,000 lb stainless NTEP shear beam load cell kit is designed for builders and service technicians who need four moisture-sealed single-ended beam sensors, mounting feet and spacers for a multi-point platform, floor scale, pallet scale, hopper or tank weighing system. Each cell is rated 5,000 lb, giving the four sensors 20,000 lb of combined nameplate cell capacity before deck dead load, corner loading, shock loads and structural safety factors are considered.

What Comes in the Four-Cell Kit?
MNM lists four 5,000 lb stainless steel, moisture-sealed shear beam load cells, four 1/2-20 adjustable feet and four spacers. These components form the sensing and support points for a four-corner scale. A complete scale normally also needs a rigid deck or vessel support structure, a summing/junction box, home-run cable, a compatible indicator and a calibration procedure after mechanical installation.
Published Electrical and Performance Specifications
MNM publishes 3.00 ± 0.015 mV/V rated output, ±0.02% full-scale nonlinearity, ±0.02% full-scale hysteresis, ±0.01% full-scale repeatability and ±0.02% full-scale creep over 30 minutes. Zero balance is ±1.00% full scale. Input resistance is 385 ± 5 ohms, output resistance is 350 ± 3 ohms and insulation resistance is at least 5000 megaohms. Excitation is specified at 10 V DC/AC. The compensated temperature range is -10°C to 50°C and the operating range is -20°C to 60°C. Published safe overload is 120% full scale.
Four-Wire Color Code
The MNM listing gives red as +input/excitation, black as -input/excitation, green as +output/signal and white as -output/signal. Confirm the actual cable and indicator terminals before wiring because field repairs or previously modified systems can differ. In a four-cell scale, each cell normally lands in a junction box so the four signals can be summed and corner-adjusted before one cable runs to the indicator.

How Four Shear Beams Work Under a Platform
A common floor-scale design supports the deck at four points, one load cell near each corner. When a load is placed on the platform, each cell carries part of the total. The junction box sums the millivolt signals. Corner-trim potentiometers compensate for small differences in mechanical leverage and sensor output so the same test weight gives nearly the same total reading regardless of where it is placed on the deck. This is why corner adjustment comes before final span calibration.
Do Four 5,000 lb Cells Make a 20,000 lb Scale?
Not automatically. Twenty thousand pounds is the arithmetic sum of the four cell ratings. A finished scale must also support its own deck weight, tolerate loads that are not perfectly centered, account for forklifts or rolling loads if those are expected, and keep the frame, feet, welds, stops and hardware within their engineering limits. A very concentrated corner load can overload one cell even while total scale weight is below 20,000 lb. Establish the finished system rating from the complete mechanical design rather than the sum alone.
Published Interchange Families
MNM publishes interchange references that include Artech SS30310, Celtron SQB SS, HBM B35, Revere 9123, Rice Lake RL35023S, Rice Lake RL39123, Sensortronics 65023SS and Coti Global CG-23SS. An interchange listing is a strong starting point, but a replacement job should still verify rated capacity, bolt pattern, loading direction, dimensions, cable length, output, bridge resistance and regulatory requirements before installation.

Installation Checklist
Mount each cell in the designed orientation with the fixed end rigidly attached and the live end free to deflect through the intended foot or loading hardware. Level the platform so all four supports share load. Prevent the deck from binding against ramps, walls, conduit or debris. Route cables away from pinch points and welding current. Before calibration, place a test load over each corner and look for a dramatically different response, which can indicate a reversed signal wire, mechanical bind, damaged cell or incorrect mounting.
Calibration Sequence
A practical sequence is: level and free the deck mechanically; verify all four cells respond in the same direction; perform corner tests and adjust the junction box; establish unloaded zero; apply a known span load; then verify several loads and multiple deck positions. The dedicated calibration article linked below goes deeper into corner trim and the difference between correcting a corner error and changing overall span.
Common Buyer Questions
Can these cells be used outdoors? The stainless, moisture-sealed construction is intended for demanding environments, but the full installation still needs protected junction-box connections, cable routing and appropriate washdown/weather practices. Are they Legal for Trade? The MNM listing identifies the kit as NTEP; the complete scale and installation must still meet applicable configuration, sealing and local certification requirements for commercial transactions. Can one damaged cell be replaced? Often yes, but verify the replacement specification and recalibrate/corner-trim the system afterward.