A&D FC-2000i fine mode allows average unit-weight calculations down to 0.002 g while the direct weighing display remains 0.2 g. Normal mode uses a 0.04 g minimum unit weight. The difference is possible because parts counting estimates the average from a group of pieces; it does not turn the scale into a 0.002 g precision balance.

When fine mode helps

Fine mode is useful when many lightweight pieces can be sampled together. Increase sample size as unit mass falls and allow ACAI to refine the average. If count accuracy remains inconsistent, check whether parts actually have uniform mass. A counting scale cannot mathematically eliminate real variation in the product.

2 kg calibration

A&D specifies a 2 kg ±0.05 g calibration mass. Level and stabilize the scale, complete digital calibration and verify with an independent known load. Keep calibration and count troubleshooting separate: a scale can pass weight verification while giving poor counts because the stored unit weight or sample is wrong.

Serial and relay options

RS-232C is standard. FC-03i adds two serial channels for multi-device connections; FC-04i adds one serial channel plus comparator relay. They cannot be installed together. USB/Ethernet/Bluetooth workflows can be built through A&D serial converters where supported, allowing the scale to participate in inventory or quality systems.

FC-05i remote platform

Connect a larger external scale with FC-05i for bulk quantities. Use the local scale to establish an appropriate unit weight, then move to the remote platform for the production lot. Verify the hand-counted result before trusting a new two-scale configuration, especially for very light pieces.

Frequently asked questions

Normal minimum unit weight? 0.04 g. Fine mode? 0.002 g. Direct readability? 0.2 g. Calibration mass? 2 kg. Is standard RS-232C included? Yes. Can FC-03i and FC-04i coexist? No. Can it use a remote scale? Yes with FC-05i. Why might count drift after supplier change? Average piece mass may have changed.

 
 
 