How to Calibrate, Verify & Troubleshoot the A&D EK-2000i

Jul 31
3 min read

Need more details or help choosing the correct item? Call MNM Scales at (832) 290-3120 or email mnmscales@yahoo.com.

A&D EK-2000i calibration uses an external reference mass, with MNM listing 2,000 g as the default setting. The model’s 0.1 g repeatability and ±0.1 g linearity specifications mean a good maintenance routine should separate calibration errors from installation, temperature, vibration, static and pan-contact problems. Because EK-2000i is non-NTEP, users can manage calibration according to their internal quality system without the trade-scale sealing rules that apply to certified EK-i models.

Prepare the balance before calibration

Level the balance with the bubble and adjustable feet, place it on a rigid bench and ensure the 133 × 170 mm pan touches only its weighing support. Avoid fans, HVAC outlets, direct sunlight and nearby vibration. Let a balance or calibration mass moved from a different temperature acclimate before adjustment. Many apparent calibration problems disappear when the installation becomes stable.

External calibration with 2000 g

Use an appropriate known 2,000 g mass and follow the EK-i external calibration procedure. Start with an empty stable pan, center the reference carefully and avoid leaning on the bench while the balance captures the value. If your quality program uses a user-definable mass permitted by the instrument, document that setting. An unverified shop object is not an adequate calibration reference when traceability matters.

Verification after calibration

Remove the calibration mass, confirm zero and apply another known value. For processes spanning a wide range, check low, middle and high working points. Repeat a key mass several times. If the spread is larger than expected, improve the environment before adjusting span again. Calibration changes the instrument; verification provides evidence that the complete setup is suitable for the process.

Repeatability and linearity checks

MNM lists 0.1 g repeatability and ±0.1 g linearity. A repeatability test uses the same mass several times with complete unload/zero between readings. A linearity check uses different known masses across the range. If one region of the range is consistently wrong while another is correct, repeating a single full-span calibration may not fix the underlying issue; confirm reference weights and arrange service if a stable load-dependent error remains.

Zero drift, static and temperature

If zero wanders, inspect for debris beneath the pan, surrounding contact, a flexible bench, airflow or changing room temperature. Plastic containers, film and powder can create static that pulls on the load. If the reading changes when a hand approaches a charged container, use static-control methods rather than recalibrating around the force. Warm or cold samples can also cause convection until they approach room temperature.

Eccentric loading and pan fit

Compare the same known weight in the center and representative off-center positions if the real application uses wide containers or inconsistent placement. Large position-dependent errors point to leveling, pan seating, mechanical contact or damage. A span calibration taken in the center cannot correct a tray that touches the housing or bench at one position.

RS-232 faults are not calibration faults

If the local display passes known-weight verification but the computer receives nothing or garbled values, troubleshoot cable pinout, baud/parity/data settings and output mode. Do not recalibrate the balance to solve a serial problem. If both local and transmitted results fail the same known-weight check, then investigate the weighing system.

Frequently asked questions

What calibration mass does MNM list for EK-2000i? 2,000 g. Should calibration be followed by verification? Yes. What are repeatability and linearity? 0.1 g and ±0.1 g. What commonly causes zero drift? Debris, vibration, static, temperature or mechanical contact. What if different pan positions produce different values? Check eccentric loading and pan contact. What if the PC reading is wrong but the display is right? Troubleshoot RS-232C. Does successful calibration make EK-2000i legal for trade? No; it remains a non-NTEP model.

 
 
 