LP7515 Animal Weighing & Peak Hold Guide: Stable Livestock and Force Readings
- pete8781
- Aug 5
- 3 min read
Need help matching the LP7515 to a livestock platform or force-testing system? Call MNM Scales at (832) 290-3120 or email mnmscales@yahoo.com.
LP7515 animal weighing depends on selecting the right hold mode. A moving cow, hog or dog creates a changing signal that should be averaged; a compressed spring or pulling fixture may need the highest force captured; a stable pallet may need no hold at all. The LP7515/OP-902 family provides peak, manual, automatic, average-animal and auto-average-animal modes, and each solves a different measurement problem.

Choose the Correct Hold Mode
Peak hold captures the highest measured value and is useful for spring compression, actuator thrust, pull-force tests and other controlled force applications. It is usually the wrong choice for live animals because one hoof impact or gate contact can create a brief high reading that does not represent body weight. Manual hold freezes a reading when the operator presses the control. Automatic hold waits for a sufficiently stable value and captures it without a separate button press.
Average hold and auto-average hold are designed for motion. The indicator samples changing weight over a selected averaging period and displays a representative result. Longer averaging can calm a restless-animal signal but also slows the workflow; shorter averaging is faster but may be less repeatable. The OP-902 procedure allows an averaging time from one to nine seconds. Auto-average begins after the load rises above its activation threshold, making it useful when animals enter and leave a platform repeatedly.
Livestock Weighing Workflow
Start with an empty, clean platform and verify zero. Keep gates, rails, chute panels, manure buildup and surrounding concrete from touching the live weighing deck. Lead the animal fully onto the platform, close the gate without letting it bear on the deck, and wait through the complete averaging interval. Record the held value, release the hold as required, remove the animal and confirm that the platform returns to zero before the next weighing.
For herd-management consistency, use the same platform, gate position, averaging time and weighing routine each session. Animal weight can vary with feed, water and timing, so process consistency matters alongside scale accuracy. The LP7515 is suited to non-NTEP livestock and process applications; legal-for-trade transactions require a complete approved system and local compliance, not merely an indicator with fine displayed resolution.

Using Peak Hold for Springs and Force Testing
For a spring tester, align the force through the load cell’s intended axis and use a fixture that cannot eject the spring. Zero the unloaded fixture, reset peak hold, apply force smoothly and stop before any system component reaches its rated limit. Peak hold records the maximum encountered value; it does not prove that the force was applied at the correct displacement. Spring-rate testing therefore needs a separate, repeatable displacement measurement and several force-versus-travel points rather than one peak number.
Why Animal Readings Stay Unstable
Persistent motion after averaging often comes from more than the animal. Check for a rocking platform, an uneven foot, a chute panel touching the deck, a cable being pulled, moisture in the junction box, one load cell carrying little weight, or a damaged connector. Compare readings at all corners with a known load. If one location differs, correct the mounting and trim balance before changing indicator filters. If the display moves negative under load or jumps when a cable is flexed, inspect wiring before recalibrating.
Related Guides for the Complete System
Read the MNM livestock scale kit complete guide for platform components and the corner-trim and calibration guide when readings vary by location. Force-test users can continue to the 1,000 lb race-car coil spring guide. For signal problems, use the LP7515 wiring guide; for zero and span, use the LP7515 calibration guide.